In most industrial and water and wastewater plants, phosphorus measurements require that a technician pull a sample manually and take it to a lab for analysis. The phosphorus measurement is performed on line with this instrument, as accurate as a lab analyser, and much less time consuming.
The CA80TP measures phosphorus in ranges of 0.05 to 10 mg/l or 0.5 to 50 mg/l. It has two 4-20mA outputs with the option for additional outputs including Modbus RS485, Modbus TCP and EtherNet/IP. The 4-20mA outputs transmit the phosphorus measurement value. The other outputs can be used to transmit the process variable, plus status and diagnostic information.
Phosphorus occurs as phosphate in natural water systems and in wastewater. Phosphates
enter the water from fertilizers, biological and industrial waste and wastewater—and from substances added in water treatment for corrosion protection. Phosphate is a nutrient, so over-enrichment of phosphate (eutrophication) leads to excessive growth of aquatic plants. When these plants die in the fall, the decay of the additional biomass increases the rate of oxygen consumption. In extreme cases, this may lead to fish kills and decrease the quality of the water system.
To determine total phosphorus, a sample must be digested. Samples are drawn from the process using a suction strainer or can be taken from a bypass piping system using an Endress+Hauser Y Strainer.
A precise sample volume is digested at high temperature with a digestion reagent. An optical dosing unit, used to dose sample and reagents, uses a redundant light barrier for a high level of reliability, guaranteeing precise and reproducible results from extremely small volumes of reagent. The temperature in the reaction vessel is kept constant to ensure complete digestion takes place within a short period of time. The CA80TP has a software-controlled safety cover to prevent access to the digestion reactor when it is under operation.
Following digestion, the sample is mixed with colorimetric analysis reagents, causing a characteristic change in the sample color. A photometer measures the level of absorption by the sample at defined wavelengths.
The analyzer can be equipped with an energy-efficient cooling system to maintain reagent quality and increase lifespan. Automatic cleaning and calibration functions ensure the analyzer works reliably and without manual intervention over an extended period of time.
If disturbances should occur, advanced diagnostics with remote access help plant personnel analyze and remedy any issues quickly. Required maintenance measures can be carried out easily, increasing process uptime while reducing maintenance costs.
@Endress_Hauser #PAuto
No comments:
Post a Comment